Main -> Products -> HPGe Detectors & Spectrometers -> Waste Assay Monitor
Waste Assay Monitor*
APPLICATION
 
The WAM-200 measuring facility is intended for the measurement and the determination of activities, activity concentrations, total activities and total activity concentrations of the selected radionuclides which emit gamma radiation in a range from 100 to 1500 keV. Total activities are the sum of activities of individual radionuclides; and total activity concentrations are the sum of all activity concentrations of individual radionuclides. Solids and the subjects with an average density up to 2500 kg/m3 located in the standard drums with a volume of about 0.2 m3 are measured.
 
FEATURES
 
The WAM (Waste Assay Monitor) is a complex measuring system which is intended for the monitoring of radioactive waste in the standard 200-litre drums. WAM involves the following systems:
  • Monitor – a fixed segmented gamma-spectrometric monitor for the determination of activities of selected radionuclides in the individual drum segments with vertical motion and collimator
  • Transfer system is used for moving the measuring part from/to the drum measured
  • MDG-12S dose rate monitor, direction-dependent, measures the dose rate of the segment in the defined distance from the drum
  • MDG-02 dose rate monitor measures the background dose rate
  • Rotary table RS01 and RS02 control and power supply switchboards.
TECHNICAL CHARACTERISTICS
 
WAM-200
Power supply
220 V +22/-33 V, 50 Hz., 2 kW, Maximum power interruption up to 20 ms. TN-S power supply system.
Communication interface
X2X, CAN, RS-485
Protection from external influence
IP 54
Material density
up to 2500 kg/m3
Material volume
up to 0.2 m3
Material weight
up to 700 kg (Maximum drum weight)
Drum measured
Type I – average of 600 mm, height of 860 mm, wall thickness from 1.4 to 1.5 mm
Type II - average of 598 mm, height of 805 mm, wall thickness of 4 mm
Dimensions
LxBxH 2500x700x2200
Weight
1700 kg
Detector
Detector
semiconductor HPGe, efficiency of 30% *
Energy range
100 to 1500 keV (resolution >2 keV at 1332 keV)
Measuring range
from minimum detectable activity of 3.7 kBq (for background of 300 nSv/hr, for 60Co in 0.2 m3 drum with an average material density of 300 kg/m3 and measuring time of 30 min.) up to: 1 GBq
Radionuclides measured
Cs-134, Cs-137, Co-60, Mn-54, Fe-59, Nb-95, Zn-65, Zr-95, Co-58, Cr-51, Ce-144, Hf-181 and Ru-103 **
Measurement precision
± 20 % (for the uniform activity distribution and the density in the volume measured) maximum 50 % for the material density of 1000 kg/m3 in the drum
 
* Other Detection Unit efficiency can be used.
** Nuclide list can be changed accordingly.
 
DELIVERY COMPLETENESS

The following parts are involved in the delivery:
 

Item specification

Quantity

1. WAM-200

1 pc

2. Spare part according to specification VF K0002-06-C-01-B05

1 pc

3. Tool kit for maintenance and calibration according to specification K0002-06-C-01-B06

1 pc

4. Instruction for assembly VF K0002-06-B-N02

1 pc

5. Instruction for use VF K0002-06-B-N01

1 pc

6. WAS Control Application S0819-01-B-N01

1 pc

 
For additional information, please contact us: sales@bsi.lv.
 
* Produced in cooperation with VF Company, Czech Republic.
Ganibu dambis 26, P.O.Box 33, LV-1005, Riga, Latvia
Tel.(+371)67383947, Fax (+371)67382620