APPLICATION
The WAM-200 measuring facility is intended for the measurement and the determination of activities, activity concentrations, total activities and total activity concentrations of the selected radionuclides which emit gamma radiation in a range from 100 to 1500 keV. Total activities are the sum of activities of individual radionuclides; and total activity concentrations are the sum of all activity concentrations of individual radionuclides. Solids and the subjects with an average density up to 2500 kg/m3 located in the standard drums with a volume of about 0.2 m3 are measured.
FEATURES
The WAM (Waste Assay Monitor) is a complex measuring system which is intended for the monitoring of radioactive waste in the standard 200-litre drums. WAM involves the following systems:
TECHNICAL CHARACTERISTICS
|
WAM-200 |
|
Power supply |
220 V +22/-33 V, 50 Hz., 2 kW, Maximum power interruption up to 20 ms. TN-S power supply system. |
|
Communication interface |
X2X, CAN, RS-485 |
|
Protection from external influence |
IP 54 |
|
Material density |
up to 2500 kg/m3 |
|
Material volume |
up to 0.2 m3 |
|
Material weight |
up to 700 kg (Maximum drum weight) |
|
Drum measured |
Type I – average of 600 mm, height of 860 mm, wall thickness from 1.4 to 1.5 mm
Type II - average of 598 mm, height of 805 mm, wall thickness of 4 mm |
|
Dimensions |
LxBxH 2500x700x2200 |
|
Weight |
1700 kg |
|
Detector |
|
Detector |
semiconductor HPGe, efficiency of 30% * |
|
Energy range |
100 to 1500 keV (resolution >2 keV at 1332 keV) |
|
Measuring range |
from minimum detectable activity of 3.7 kBq (for background of 300 nSv/hr, for 60Co in 0.2 m3 drum with an average material density of 300 kg/m3 and measuring time of 30 min.) up to: 1 GBq |
|
Radionuclides measured |
Cs-134, Cs-137, Co-60, Mn-54, Fe-59, Nb-95, Zn-65, Zr-95, Co-58, Cr-51, Ce-144, Hf-181 and Ru-103 ** |
|
Measurement precision |
± 20 % (for the uniform activity distribution and the density in the volume measured) maximum 50 % for the material density of 1000 kg/m3 in the drum |
* Other Detection Unit efficiency can be used.
** Nuclide list can be changed accordingly.
DELIVERY COMPLETENESS
The following parts are involved in the delivery:
|
Item specification |
Quantity |
| 1. |
WAM-200 |
1 pc |
| 2. |
Spare part according to specification VF K0002-06-C-01-B05 |
1 pc |
| 3. |
Tool kit for maintenance and calibration according to specification K0002-06-C-01-B06 |
1 pc |
| 4. |
Instruction for assembly VF K0002-06-B-N02 |
1 pc |
| 5. |
Instruction for use VF K0002-06-B-N01 |
1 pc |
| 6. |
WAS Control Application S0819-01-B-N01 |
1 pc |
For additional information, please contact us: sales@bsi.lv.
* Produced in cooperation with VF Company, Czech Republic.