APPLICATION
The gamma–ray Detection Device is based on a HPGe detector with Electric Machine Cooling which is intended for the conversion of Gamma-ray quantum energy using amplitude electric signals and their amplification for further registration with a nuclear physics apparatus. It was specifically developed for long, continuous operations.
FEATURES
-
No liquid nitrogen necessary;
-
Detection of radiation possible in any spatial orientation;
-
Automatic restart after power supply switch-off;
-
Long-duration continuous function;
-
Coaxial or planar HP(Ge) detector can be used.
COMPLETE SET
-
Detection device consisting of cryostat, HP(Ge) detector crystal, temperature sensor and heat exchanger;
-
Gas compressor cooling system with built-in cryocontroller for the provision of automatic monitoring and control of operating modes for HP(Ge) detector and cryosystem;
-
High pressure gas pipes for connecting the cryosystem to the Detection Unit heat exchanger;
-
-
Emulation and analysis Software.
TECHNICAL CHARACTERISTICS
|
Parameter |
Value |
|
Energy range, keV |
40 - 3000 |
|
HP(Ge) detector efficiency, % |
30 * |
|
Energy resolution for 30% efficiency detector, keV
at energy
122 keV
1.33 MeV |
0.9
1.9 |
|
Deterioration of energy resolution at 662 keV line as compared with resolution of detector cooled by liquid nitrogen, % |
< 5 |
|
Compressor continuous operation time, h |
50000 |
|
Overall dimension, mm
Detector capsule
Detector capsule with cryocooler
Compressor
Spectrometric device Multispectrum |
Ø90x130
Ø114x350
445x357x281
300x180x80 |
|
Weight, kg
Detector with cryocooler
Compressor
Spectrometric device Multispectrum |
2.9
31.8
2.8 |
|
Max distance between detector and compressor, m |
15 |
|
Consumed Power, W |
570 |
|
Voltage, V |
220 |
|
Frequency, Hz |
50 |
* 10% - 160% efficient HP(Ge) detectors are available by separate order