APPLICATION
 
The Digital Spectrometric Device DigiSpectrum is a high - precision, ultra-fast all - digital spectrometer, comprising a single processing channel, a preamplifier power supply and a detector bias supply in a compact package. The Digital Spectrometric Device can accept signals from practically any radiation detector. The Digital Spectrometric Device has built-in support for HPGe detectors with a Compton shield.
 
FEATURES
  • Pulse heights measured with up to 16 bits accuracy (i.e. 32K spectrum length).
  • Programmable gain, input offset, and peaking times between 0.25 and 40 microseconds.
  • Programmable pileup inspection criteria include trigger filter parameters, threshold, and rejection criteria.
  • Peak stability with rate: 0.05% up to highest ICR.
  • Integral Nonlinearity of Energy: = 0.1% of full scale.
TECHNICAL CHARACTERISTICS
 
Signal Input (Analog) - Works with common reset or resistor feedback preamps of either polarity.
  • Impedance: 50 Ohms, 90 Ohms, 250 Ohms, and 1k Ohms - jumper selectable.
  • Attenuation: 1:21, 1:12, 1:5 and 1:1- jumper selectable.
  • Voltage Range: +10.0 V to -10.0 V standard.
Inputs (Digital)
  • Gate Input: Analog input for Compton suppression. Can alternatively be configured as digital I/O.
  • HV Inhibit: TTL logic input. BNC connector, jumper set active HI or LO.
Interface:
  • USB: Universal Serial Bus.
Digital Controls:
  • Gain: 50X coarse gain (2% accuracy), fine gain controlled by 16 bit DAC.
  • Shaping: Triangular/Trapezoidal. Peaking time and dwell set independently : 50ns - 45 microseconds in small steps. Adjustable dwell time may be used to eliminate ballistic deficit effects.
  • Pileup: Set fast channel filter time, pulse detection threshold, and fast channel pileup inspection test limits to achieve best results in each situation.
  • Spectrum: 1K-32K bins, 32 bits deep (4.3 billion counts per bin)
Collection:
  • Livetime; Realtime; input/output count rates; Compton coincidence rate.
Power Outputs:
  • Detector Supply: High Voltage +/- 5kV, SHV connector, push-button on/off, front panel adjust, 60 sec. on/off ramp standard.
  • Preamp Supply: standard preamp power voltages +/- 24V and +/- 12V, each to 100 mA.
  • Power Requirements: 110 V at 0.2 A 50/60 Hz or 220 V at 0.1 A 50/60 Hz, specify at time of order.
Software package
 
SpectraLineGP spectra processing includes calibration, peaks parameters determination, nuclides identification, activities calculation and using the true-coincident factors for the gamma-emission intensity correction. Non-parametric model for pattern of the full energy peak provides a correct model for a line in any energy range.
 
Functionalities
  • automatic peaks search with the required sensitivity level;
  • the energy, FWHM and peak pattern calibrations;
  • peaks parameters determination - position, FWHM, area; calculation results can be saved to a text file;
  • efficiency “curves” can be obtain with the efficiency calibration;
  • activity calculation by various methods;
  • use the true-coincident factors to correct the gamma-emission intensity;
  • the measured spectra and processing results saving to database for the analysis of the specified criterions convergence of the repeated measurements;
  • simultaneous processing of any spectra number;
  • the peak pattern calibration with several spectrum peaks in different energy ranges;
  • numeric and visual control of the calibrations results;
  • any measurement tracts number;
  • independent control of all channels - start, stop, etc.